59
ELECTROMAGNETIC
COMPATIBILITY
59
ELECTROMAGNETIC
COMPATIBILITY
Electromagnetic compatibility
&ODVVLF/LQHEDWKWXELVLQWHQGHGIRUWKHXVHLQWKHHOHFWURPDJQHWLFHQYLURQPHQWVSHFL¿HGEHORZ7KH
customer or the user of Classic Line 2 bathtub should assure that it is used in such an environment.
The product has been tested for compliance with current regulatory standards regarding its capacity to
block EMI (electromagnetic interference) from external sources.
Some procedures can help reduce electromagnetic interferences:
• Use only Arjo cables and spare parts to avoid increased emissions or decreased immunity which
can compromise the correct functioning of the equipment.
• Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted
emissions standards.
Intended Environment: Professional Healthcare facility environment.
Exceptions: HF Surgical Equipment and the RF Shielded room of an ME SYSTEM for magnetic
resonance imaging.
WARNING
Wireless communications equipment such as wireless computer network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies, etc. can affect this
equipment and should be kept at least 1.5 m away from the equipment.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and
the other equipment should be observed to verify that they are operating normally.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSION (EMI)
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
This equipment uses RF energy only for its internal functions.
Therefore its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
This equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low voltage power supply network that supplies buildings used for
domestic purposes.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
9ROWDJHÀXFWXDWLRQV
ÀLFNHUHPLVVLRQV
IEC 61000-3-3
Complies
Содержание Classic Line 2
Страница 1: ...INSTRUCTIONS FOR USE Classic Line 2 04 ACL 00_1EN 08 2020 ...
Страница 6: ...Intentionally left blank ...