SOFTWARE UPDATES
14.147.B_QUICK100M en.SM.V1.00.doc
ARGUS Medical AG 11.08.04 / PJ
8 / 22
ARGUS Medical AG
CH-3627 Heimberg
Bootloader V0.21
20030328 (C) AM
Loader for flashing ARGUS Medical devices.
1. Erase user program.
2. Update user program.
3. Erase bootloader.
4. Update bootloader.
1.
Unplug
the mains power cord
from the QUICK
®
100 M docking station
to reset it
and remove all pumps
.
2. Connect the device to your PC.
3. Start the terminal program and make the settings as described in chap-
ter 4.1.4.
4. Use the menu "Send text file..." in the menu "Transfer" of the Hyper
Terminal program to send the dongle file "AFlash_Dongle.txt" to the
QUICK
®
100 M docking station (this file will send synchronisation char-
acters for approx. 5 seconds).
5. After the start of the transmission, within 5 seconds
plug
in the mains
power cord
to the QUICK
®
100 M docking station.
6. Wait several seconds until the terminal program displays the following
text (if the device entered the boot loader mode correctly):
7. Follow the steps described in the next chapters based on the kind of
update you want to perform.