3. CONFIGURATIONS
3.1. Configuration of the docking station
A special software tool for the configuration of the docking station will be released
soon. Please contact your local distributor or ARGUS Medical AG.
3.2. Pump configuration
CAUTION!
Every pump used with the QUICK
®
100 M docking station has to be preconfigured with
its corresponding serial number!
Thus the addresses #396 and #397 in the configuration of each pump must be pro-
grammed with the corresponding serial number. Please refer to the pumps respective
service manual to change these addresses if necessary. In the case of an invalid serial
number, the docking station will generate an alarm and the pump can not be pro-
grammed via barcode.