SOFTWARE UPDATES
14.147.B_QUICK100M en.SM.V1.00.doc
ARGUS Medical AG 11.08.04 / PJ
10 / 22
4. After a successful download, the QUICK
®
100 M docking station will acknowl-
edge the end by ringing the internal buzzer twice.
5. Type a "5" in the main menu to quit the boot loader mode of the QUICK
®
100 M
docking station. You will be asked if you want to quit. After confirmation the
QUICK
®
100 M docking station will now run in the normal mode.
Important Remark:
Updating the boot loader program will not have any influence on the user program.
Do not unplug the mains of the device during updating the boot loader or quit this
mode without any valid boot loader in the flash memory. The device must contain
a valid boot loader to operate correctly. If it is not the case, it will no longer be pos-
sible to enter the flash mode. In this case the main board M has to be replaced!
4.4. Important Notes
The user program performs an internal ROM-Test during the self test phase of the
docking station. If an error occurred during the update process the device will en-
ter the fail safe state (fault codes please refer to chapter 7.2).
Please refer to your local distributor or ARGUS Medical AG to determine the latest
software release able to run on your device hardware.