SAFETY STANDARD CHECK
14.147.B_QUICK100M en.SM.V1.00.doc
ARGUS Medical AG 11.08.04 / PJ
22 / 22
11. SAFETY STANDARD CHECK
Safety standard check
QUICK® 100 M
ARGUS Medical AG
Serial-No.:
Hospital/Dept/Customer:
The safety standard check has to be performed at least every 24 months or after 10000 hours of operation.
The check has to be done in accordance to the operation and service manual.
1 Visual check for damage, cleanness and completeness:
- Housing, labels, accessories, connectors, cables, etc..?
2 Plug in the mains of the docking station
- The green LED on the docking station lights up?
3 Check each docking position with a switched
off
pump:
- Mechanical locking and unlocking ok ?
- After docking the pump, its green power LED lights up?
4 Check each docking position with a switched
on
pump:
- Buzzer sounds twice, ok?
- LED on docking station flashes orange-red, ok?
5 Chech the external connections:
- Computer interface ?
- Barcode reader interface ?
- Staff alerting system ?
6 Electrical test according to EN60601-1:
- Measurements attached?
Do not install any pump!
All measurements made with a 2.5 m power cord