5. HISTORY
5.1. General
The connection of the QUICK
®
100 M docking station to a computer allows to read
the history file stored in the flash memory.
5.2. Entering the docking station history read mode
To enter the history read mode please follow the procedure listed below:
1.
Switch off all docked in pumps
.
Unplug the mains power cord
from
the QUICK
®
100 M docking station and plug in the mains power cord
again to reset it.
2. Connect the device to your PC.
3. Start the terminal program and make the settings as described in chap-
ter 4.1.4.
4. Type "QUI;100" + ENTER. The history read mode program will be acti-
vated and the red LED lights up.
5. Type "HIS;" + Enter. The history will now be displayed on the screen
(see examples below).
6. Exit the history read mode by unplugging the mains power cord again.
The events registered will be displayed in a chronological order; the most recent
entry is displayed on top of the list. Please be aware that the time displayed in
minutes refers always to the elapsed time since last power up of the QUICK
®
100
M docking station.
5.3. History header
At the top of the history, a header will appear showing the device type, the soft-
ware release and if configured, the inventory number and the last service date.
5.4. History events
Each registered event starts with a message line. See the list mentioned in chap-
ter 5.6.
In case the docking station detects a technical failure, the last event line registered
includes the failure number (fault code) which generated that technical failure.
Please refer to chapter 7.2 for the complete list.