14.148.- / A414D en / Software 1.30
14.10.2003
7
1.5
List of the interrogation- and configuration functions:
CAUTION! Before you make a new configuration or replace an EPROM or the mainboard make a note of
the programmed values. Afterwards you can re-enter the old not writeprotected values.
Address Write Default
Function
(#)
prot.
Value
0
0 = No
PIN code active
1
1 = Yes
Run indication by running decimal point
2
0 = No
Key "ON/OFF" only at STOP valid
3
0 = No
Rate change only at STOP valid
4
0 = No
Key "STOP" delayed (time at #361)
5
0 = No
2nd entry of rate (#3=1), rate calculation disabled.
6
0 = No
Static alarm (staff alerting system)
7
0 = No
Display elapsed time in run mode (#8=0)
****
8
0 = No
Select remaining time (#7=1)
9
1 = Yes
Alternative time input (10h, 1h, 1/4h)
10
0 = No
Operation without drop detector
11
1 = Yes
Recall "ml/h" (rate) at next power on
12
0 = No
Recall "ml total" (end volume) at next power on
13
0 = No
Recall "ml inf." (volume inf.) at next power on
14
0 = No
SBS Step By Step function
15
0 = No
Display VTBI (Volume To Be Infused)
16
0 = No
Display "SEt -X-"’ if only 1 set enabled
17
1 = Yes
KVO (KOR), mode see #60
18
0 = No
Drop alarm only if bottle is empty (#10=0)
19
1 = Yes
Buzzer at start
20
0 = No
Menu "CLr" (clear "ml inf.") (#15=0)
****
21
0 = No
Menu "uOP" (Micro operation)
****
22
0 = No
Menu "trA" (transport) (#10=0 and #18=0)
****
23
1 = Yes
Menu "PrL" (pressure alarm limit)
****
24
1 = Yes
Menu "CAP" (battery capacity)
****
25
0 = No
Menu "SEt Fill"
****
26
0 = No
Menu "InF" ("ml inf." since last power on)
****
27
0 = No
Menu "dLo" (data-lock)
****
28
0 = No
Menu "Stb" (stand-by)
****
29
0 = No
Menu "MEd" (medication number)
****
30
0 = No
Menu "tM " (timer alarm)
****
31
0 = No
32
0 = No
Menu "boL" (release bolus)
****
33
0 = No
Menu "boLr" (bolus rate) (#32=1)
****
34
0 = No
Menu "tot" (bolus total) (#32=1)
****
35
0 = No
36
0 = No
37
0 = No
38
0 = No
Bolus application automatic (#32=1 and #34=1)
39
0 = No
40
0 = No
Demo mode (all menus enabled)
****
41
0 = No
Clear "ml/h" after infusion completed
42
0 = No
Clear "ml total" after infusion completed (#41=1)
43
0 = No
Air volume accumulated (1ml over 0.5 hrs)
44
1 = Yes
Automatic pressure release after occlusion
45
1 = Yes
Pressure display 20/40/60/80/100% (Bargraph ON)
46
0 = No
Bargraph with indicator (25% steps, #45=1)
47
0 = No
Standby- and battery prealarm low volume
48
1 = Yes
Flashing numeric display at alarm
49
0 = No
Alarm acknowledge only with key "MODE"