12
14.10.2003
14.148.- / A414D en / Software 1.30
1.6
Special configuration options
- Configuration of a reminder alarm for the safety standard check:
First the service interval has to be configured either in months or in hours of operation, or both (addresses
393, 394).
Next the last service date has to be entered on address 390. Any value higher than 0 entered at the adresses
393 and/or 394 will release the reminder alarm after the set service interval has elapsed (check also the
correct settings of the internal clock).
- PC configuration tool "AConfig":
With this additional software the pump may be configured from a PC over the serial port. This software
may be available from your local distributor or our service department.
- IV-set definition:
Allows the individual calibration of up to 4 dedicated IV-sets over the full infusion rate range. Please
contact your local distributor or our service department for further information.
After changing the configuration a function check and a control measurement has to be made!