![Argus Green Stream VO-P ARGUS 414 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/argus/green-stream-vo-p-argus-414/green-stream-vo-p-argus-414_service-manual_2969443013.webp)
14.148.- / A414D en / Software 1.30
14.10.2003
13
2
History and pump configuration printout
2.1
Connecting of the ARGUS414 to the serial interface
Caution: The infusion pump must be disconnected from the patient before any connection
over the serial interface can be done.
A connection of the ARGUS414 to a computer is useful to the read the actual configuration or history of the
pump. Even a simple monitoring of the pump can be done over the serial interface RS-232.
The connection of the infusion pump with your computer over the interface can be done by connecting the
interface cable and adapter (part 12.011 and part 12.012) and the following steps:
-
Connect the RS-232 interface cable to outlet (32) of the infusion pump and to the serial port of your PC.
Note in which port (COM1 or COM2) you have pluged in.
-
Start your terminal program on your computer. A simple terminal program, e.g. "Hyper Terminal" is
included in every MS-Windows 9x and Windows NT systems, but must be installed first if necessary.
-
Be sure that you have selected the right serial port (COM1 or COM2) and set the following communica-
tion parameters:
Bits per second: 4800 Baud
Data bits:
8 bits
Parity:
None
Stop bits:
1 bit
Protocol:
None
-
Go to the next step in one of the further chapters, depending on your intention.
2.2
Configuration printout
-
Switch the pump on while keeping the keys "MODE" and "START/STOP" pressed and go in the configura-
tion mode.
-
Select address 399 on the left hand display.
-
Start capturing text received over the serial interface, e.g. by selecting "capturing text..." in the menu of
the Hyper terminal. A text file which contains the actual configuration printout will now be generated.
-
Enter the data 3456 on the right display of the pump.
-
Press the "START/STOP" key.
-
The pump will now transfer the actual configuration of the pump in the format mentioned below.
-
Stop the capturing of the text received; this will also close the text file generated.
-
The generated text file can be opened and printed out by any text program.
Pump configuration printout (sample)
/***** Configuration profile *****/
Wed 19-Jan-
2002
11:29:55
Pump type
: ARGUS414
Inventory number
: 0000 0000
Software release
: V1.00 (000719-4061)
Infused sum
: 678ml
Operating time
: 5h32min
Last service date
: 2000 week 12
00=0
50=0
100=0
300=0
320=0400
340=0000
360=0000
380=0000
01=1
51=0
101=0
301=0
321=0000
341=0000
361=0000
381=0000
02=0
52=0
102=0
302=0
322=0000
342=0000
362=0000
382=0000
03=0
53=1
103=1
303=0
323=0000
343=0000
363=0000
383=0000
etc.