Replacement of parts
4.1
Disassembly of the pump
NOTE: The exploded views in the appendix show the individual mounting steps.
CAUTION!
Switch the unit off and disconnect the mains cable from the power outlet before opening the housing. The
antistatic protection have to be strictly adhered to when the ARGUS infusion pump is disassembled (the use
of an antistatic table mat and a grounded clip are recommended, for example).
a Disassembly of the hood:
Remove the four screws on the side and on the rear wall, lift hood up and disconnect the plug-type
connector of the drop barrier.
b Disassembly of the main printed board assembly:
Disconnect the plug-type connector of the main printed board assembly and unscrew the two lateral
attachment screws. Please refer to the wiring diagram in the appendix.
c Disassembly of the front panel:
Use a ball-headed hexagon screwdriver to unscrew the two attachment screws from the rear and remove
the front panel. This special screwdriver can be obtained from the service department on request.
d Disassembly of the pump unit:
By unscrewing the four attachment screws, the whole pump unit can be removed from the front panel.
e
Remove the pump door:
The pump door can be remove without any tools by pulling it out in the
direction of the arrow according to the drawing. For the dismounting and the
mounting the opening angle from 30 to 45
°
has to be kept.
4.2
Replacement of the EPROM or the display board
a
Software updates may reset automatically the default values in the EEPROM. You are requested to write
down the actual contents of the addresses before you replace the EPROM or the display board. Afterwards
you have to re-enter these values in the program mode. If a write protection code has been used before, the
same code has to be programmed again.
b
A pressure sensor calibrating is necessary when replacing the pressure sensor, a pressure sensor
calibration and a volume calibration are necessary when replacing the EPROM or the display board. Be
careful and carry out the described instructions step by step.