14.148.- / A414D en / Software 1.30
14.10.2003
17
3
Fault finding
The fault code in case of a failure is displayed by pressing "MODE" key (21). (F-XX) appears in display (9),
and the source of the failure is listed in the table below:
Error
Error reason
Assembly group
F-21...22
ROM test
Mainboard
F-23...24
RAM test
Mainboard
F-25
CPU test
Mainboard
F-26
Invalid function menu
F-27
EEPROM data invalid
Mainboard
F-28
RTC (real time clock) data invalid
Mainboard
F-29
Stepper motor power test
Mainboard
F-31
1.2 Volt supply out of range
Mainboard
F-32
5 Volt supply out of range
Mainboard
F-33
30 Volt supply out of range
Mainboard
F-41
Pressure sensor test failed
Pressure sensor or mainboard
F-42
Air detector test failed
Air detector or mainboard
F-43
Air bubble size invalid
Mainboard
F-44...45
Address invalid for EEPROM
Mainboard
F-46
Frequency from
µ
C or RTC out of range
Mainboard
F-47
Displayboard not present
F-48
Key(s) too long active
Displayboard
F-49
Set correction invalid
Mainboard
F-50
Pressure monitor
Mainboard or pressure sensor
F-51...56
Rotation control
- Mainboard
- Hallboard
- Motor
- Clutch
F-57...59
Volume control
Mainboard
Exceptionally a fault code may appear, which is not included in this list. In this case we recommend to
change the main board.