Volume calibration general
1
Go into the program mode.
2
Decide which IV set should be calibrate and check its release address.
- address 101 for IV set no. -1-
- address 102 for IV set no. -2-
- address 103 for IV set no. -3-
- address 104 for IV set no. -4-
Set the correction value in one of the following address to 1000:
- address 319 for IV set no. -1-
- address 329 for IV set no. -2-
- address 339 for IV set no. -3-
- address 349 for IV set no. -4-
Switch the pump off and on.
3
Select the IV set you like to calibrate and make the following pump settings (for warm up the peristaltic
system): Rate = 999 ml/h; volume "total" = 10 ml
Start the pump by pressing "START/STOP" once. Switch the pump off and on again after the volume
"total" is reached.
4
The next pump settings are:
Rate = 100 ml/h; volume "total" = 25 ml
Start the pump by pressing "START/STOP" once. Switch the pump off immediately after the volume
"total" is reached, the net weight result must be 25 g +/-5%.
5
Calculate the correction factor with the equation:
Correction factor = (measured volume / preset volume) * 1000
6
Go into the program mode and select address for the correction value (see point 1)
Press "MODE" to enter the correction factor in the right hand display.
Press "START/STOP"] to acknowledge the correction factor.
Switch the pump off.
7
Make a control measurement with the same settings as mentioned in point 4, using the IV-set for which
the correction factor has been changed. Perform an occlusion pressure check (see chapter "Pressure
sensor calibrating", point 15-19) to verify the pressure alarm level.