
PUMP CONFIGURATIONS
14.164.A_A707en.SM.V1.01.doc
ARGUS Medical AG 10.02.04 AS
8 / 38
2.6.
Address list of the pump configuration
Add.: Address
Def.: Default value of the manufacturer
P:
Protection grade (W = Write enabled)
Add. Def. P Function
Range
0
0
W PIN code active
0=No; 1=Yes
1
1
W Run indication by running decimal point
0=No; 1=Yes
2
0
W Key [ON/OFF] only at STOP valid
0=No; 1=Yes
3
0
W Rate change only at STOP valid
0=No; 1=Yes
4
0
W Key [STOP] delayed (time at #361)
0=No; 1=Yes
5
6
0
W Static alarm (staff alerting system)
0=No; 1=Yes
7
0
W Display elapsed time in run mode (#8=0)
0=No; 1=Yes
8
0
W Select remaining time (#7=1)
0=No; 1=Yes
9
10
0
W Operation without drop detector
0=No; 1=Yes
11
1
W Recall "ml/h" (rate) at next power on
0=No; 1=Yes
12
0
W Recall "ml total" (end volume) at next power on
0=No; 1=Yes
13
0
W Recall "ml inf." (volume inf.) at next power on
0=No; 1=Yes
14
0
W SBS Step by step function
0=No; 1=Yes
15
0
W Display VTBI (Volume To Be Infused)
0=No; 1=Yes
16
17
1
W KVO (KOR, mode see (#60)
0=No; 1=Yes
18
0
W Drop alarm only if bottle empty
(#10=0, => 'trA' as default)
0=No; 1=Yes
19
1
W Buzzer at start
0=No; 1=Yes
20
0
W Menu 'CLr' (Clear "ml inf.") (#15=0)
0=No; 1=Yes
21
22
0
W Menu 'trA' (Transport)(#10=0)
0=No; 1=Yes
23
1
W Menu 'PrL' (Pressure alarm limit)
0=No; 1=Yes
24
1
W Menu 'CAP' (Battery Capacity)
0=No; 1=Yes
25
0
W Menu 'SEt Fill'
0=No; 1=Yes
26
27
0
W Menu 'dLo' (Data-lock)
0=No; 1=Yes
28
0
W Menu 'Stb' (Stand-by)
0=No; 1=Yes
29
0
W Menu 'MEd' (Medication-no.)
0=No; 1=Yes
30
0
W Menu 'tM ' (Timer alarm)
0=No; 1=Yes
32
0
W Menu 'boL ' (Release Bolus)
0=No; 1=Yes
33
0
W Menu 'boLr' (Bolus rate)(#32=1)
0=No; 1=Yes
34
0
W Menu 'tot ' (Bolus total)(#32=1)
0=No; 1=Yes
38
0
W BOLUS application automatic (#32=1,#34=1)
0=No; 1=Yes
40
0
W Demo mode (menus addr. 20..34 and 7 enabled)
0=No; 1=Yes
41
0
W Clear "ml/h" after infusion completed
0=No; 1=Yes
42
0
W Clear "ml total" after inf. completed (#41=1) 0=No; 1=Yes
43
0
W Air volume accumulated, see #350 and 351
0=No; 1=Yes
44
1
W Automatic pressure release after occlusion
0=No; 1=Yes
45
1
W Pressure display (bar graph)
0=No; 1=Yes
46
0
W Pressure display with indicator (#45=1)
0=No; 1=Yes
47
0
W Standby- and battery pre alarm low volume
0=No; 1=Yes
48
0
W Flashing numeric display at alarm
0=No; 1=Yes
49
0
W Alarm acknowledge only with key [MODE]
0=No; 1=Yes