PUMP CONFIGURATIONS
14.164.A_A707en.SM.V1.01.doc
ARGUS Medical AG 10.02.04 AS
10 / 38
Add. Def. P Function
Range
370 0
W Clock
seconds
(0...59)
371 0
W Clock
minutes
(0...59)
372 0
W Clock
hours
(0...23)
374 0
W Clock
days
(1...31)
375 0
W Clock
months
(1...12)
376 0
W Clock
years
(2000...2099)
380 0
Last failure number (F-XX)
Refer to chapter 5.7
381 0
Last infusion rate at failure
382 0
.
383 0
.
384 0
.
385 0
.
386 0
.
387 0
.
388 0
Oldest failure number (F-XX)
389 0
Oldest infusion rate at failure
390 0
W Last service date
(yyww, year and week)
391 0
2. last service date
(yyww, year and week)
392 0
3. last service date
(yyww, year and week)
393 0
W Service interval in months
(0 = disabled)
(0...24)
394 0
W Service interval in hours of
operation (0 = disabled)
(0...9999)
395 0
W Own address for SCI
(0 = no address)
(0...127)
396 xxx
W Serial number of the pump
(xxxx yyyy)
[xxxx....]
397 8yyy W Serial number of the pump
(xxxx yyyy)
[....yyyy]
399 707
Data xxxx -> clears protection key
2.7.
Special configuration options
1.
Configuration of a reminder alarm for the safety standard check:
If a safety standard check reminder alarm is configured, the pump will display
“CtrL” after power up when the time has come to perform a standard safety check.
To configure a reminder alarm please follow next steps:
First the service interval has to be entered either in months or in hours of operation,
or both (address 393, 394).
Next the last service date has to be entered on address 390. Any value higher than 0
entered at the address 393 and/or 394 will release the reminder alarm after the ser-
vice interval has elapsed (check also the correct settings of the internal clock).
2.
PC configuration tool “AConfig”:
With this additional software the pump may be configured from a PC over the serial
port. This software may be available from your local distributor or ARGUS Medi-
cal AG.
Caution: AConfig may only be used with software versions greater or equal to
1.01.