MAINTENANCE
14.164.A_A707en.SM.V1.01.doc
ARGUS Medical AG 10.02.04 AS
19 / 38
2. Close the
roller
clamp and
wait 10
seconds.
4. Press key
"MODE" to
enter the next
calibration
procedure.
The display
shows
“CAL. 0.4b”.
3. Press the
"START /
STOP" key to
store the
value for an
upstream oc-
clusion. This
will be ac-
knowledged
by a sound.
5. Open the roller clamp for the
next step!
1. Insert a new IV-set filled with water, place the roller
clamp on the upstream side near the pump and close
the door. Open the roller clamp and let the pump
running for 5 seconds at least.
5.3.4.
Calibrating the upstream sensor part 1 (left side)
Remark:
By pressing the key "1" the pump display changes to “xxxx Pr S” where the left
display shows the actual upstream pressure sensor signal in mV.
Return to the original state by pressing the key "1" again.