PECA 30 / 30 E : programming-configuration
24
ARDETEM
JbUS
brd.1
brd.2
Sor.1
brd.1
PULS
no
brd
99
JbUS
ESCL
CONFIGURATION OF OPTIONAL BOARDS
PROGRAMMING MENU :
Example
: board Nr1 = board of 3 pulse outputs board n°2 = option not available on PECA 30
board choice
board type
( details below)
board Nr1
no board
Modbus
Nr2
board
output choice
programming of the chosen output
programming of the digital
Current output
see page 31
output, Modbus / Jbus
Pulse output
see page 28
see page 34
Alarm threshold
see page 25
•
TYPE OF BOARD :
The analyser displays «
ALAr
» : 3 alarms output type board
(see p25)
The analyser displays «
PULS
» : 3 pulses output type board
(see p28)
The analyser displays «
AnA
» : 3 analog outputs type board (current)
(see p31)
The analyser displays «
AnA.A
» : 3 analog (current) o2 alarms mixed type board (see p31 & 25)
The analyser displays «
AnA.P
» : 3 analog (current) o2 pulses mixed type board (see p31 & 28)
•
INPUT / OUTPUT BOARDS :
The board number corresponds to its position inside the analyser.
The boards have 3 inputs / outputs of the same time, except the mixed board (type « AnA.A » or « AnA.P »)
which have 2 outputs in more.
Type board « ALAr » : outputs 1,2,3 = threshold relays
Type board « PULS » : outputs 1,2,3 = pulse generator (on relay contact)
Type board « AnA » : outputs 1,2,3 = current outputs 4 20, 0 20, 0 10, 0 5, -20+20, -10+10 or -5+5mA
Type board « AnA.A » : outputs 1,2,3 = current outputs 4 20, 0 20, 0 10, 0 5, -20+20, -10+10 or -5+5mA
outputs 4,5 = threshold relays
Type board « AnA.P » : outputs 1,2,3 = current outputs 4 20, 0 20, 0 10, 0 5, -20+20, -10+10 or -5+5mA
outputs 4,5 = pulse generator (on relay contact)