Model 300T2G8
To activate Sleep Mode locally:
Press the MORE soft key to get to MENU 2. At MENU 2 press the SETUP soft key to get to SETUP 1. At
SETUP 1 press SET to activate Sleep Mode (turn heater and fan off). The system will ask
Are you sure?
Press
SET again.
After activating the Sleep Mode:
Screen will display
Cooling On
while heaters cool down.
System Off
notifies user that the amplifier is in Sleep
Mode
To deactivate Sleep Mode locally:
Press the ON soft key to de-activate Sleep Mode. Amplifier will return to MENU 1. When de-activating the Sleep
Mode the heaters will require approximately a 5 minute heater time delay. Wait the full 5 minutes prior to
selecting OPERATE.
For remote activation of Sleep Mode or to set the Auto Sleep Mode timer remotely see Table 3-3 in Section 3.8.
Warnings Screens
- From MENU 3, S1 (labeled WARNINGS) selects WARNINGS 1 that allows the operator to
enter the maximum forward power. The existing value is between brackets[]; pressing SET puts arrows >< around
the value, indicating that the adjust knob is active. The effect of the warning setpoint is as follows: if the forward
power exceeds the setpoint, the audible alarm will sound (if configured in SETUP 3).
This warning will be repeated every thirty seconds until the over forward power condition is cleared. In addition,
a warning message will appear on line 3 (the status line) of MENU 1. In the event that the alarm is heard, the
operator should go to MENU 1 to determine the cause.
Pressing more brings up WARNINGS 2, which allows the under forward power setpoint to be entered. Adjusting
this to the minimum value causes -OFF- to be selected, disabling this alarm.
In WARNINGS 3, the maximum reverse power level is set. Note that these are warning levels at which the beep
sounds; the actual maximum reverse power level that generates a system fault is set in hardware in the TWT
power supply HPA Logic and Control module (A16485).
MORE brings up WARNINGS 4, which allows input of the maximum cabinet temperature. Entering this
parameter is performed as above.
MORE brings up WARNINGS 5, identical to the previous screen except that it deals with the maximum TWT
collector block temperature. If either parameter exceeds the setpoint, the audible alarm will sound every 30
seconds (if configured), and a warning message will appear on line 3 of MENU 1.
From WARNINGS 5, MORE brings up WARNINGS 6, which permits setting the maximum helix current. Any
helix current above this setpoint will result in an audible alarm (if configured), repeated every 30 seconds; and a
warning message is displayed on the status line of MENU 1.
Pressing MORE again returns display to WARNINGS 1. As before, pressing EXIT from any of the warnings
screens returns display to MENU 3.
Info Screen
- From MENU 2, S2 (labeled INFO) selects a screen that displays the RF sample port calibration
factors at various frequencies across the band. In addition, this screen displays the model number, serial number
16
Rev A
Содержание 10012514
Страница 2: ......
Страница 4: ......
Страница 14: ......
Страница 16: ......
Страница 24: ......
Страница 46: ...Model 300T2G8 28 Rev A...
Страница 50: ...Model 300T2G8 32 Rev A...
Страница 54: ...Model 300T2G8 36 Rev A...
Страница 55: ...5 3 WIRING DIAGRAMS 20 22520 000 TWT Power Supply 10 1 2 Series A22520 052 Rev A 37...
Страница 56: ...Model 300T2G8 38 Rev A...
Страница 72: ......