17. Appendix 1. Flow Through Cell
17.1. Introduction
The AP-Lite is designed to be used with the AP-2000 Flow Through Cell (Flowcell) via an
adapter collar.
The Flowcell allows sample water to flow through the AP-Lite, passing over the Electrode.
This eliminates air contact with pumped samples from groundwater boreholes allowing truly
representative measurements to be obtained.
Made from marine grade aluminium and 6mm wall thickness acrylic, the Flowcell is
ruggedly constructed for hard use in the field. The base flange includes four holes to allow
the unit to be pegged down if necessary.
17.2. Adapter Collar Installation
Flowcell Adapter Collar
1. Remove the sleeve from the AP-Lite.
2. Screw the Adaptor Collar onto the AP-Lite as shown below.
3. Re-fit the Sleeve as shown below. Be sure to also fit the Sleeve End Cap.
4. Carry out a full calibration with the Adaptor fitted.
Re-calibration at this point is extremely important because by fitting the Adaptor Collar, you
have altered the dimensions of the measurement chamber and hence the calibration.