6. Connecting the AP-Lite to an Aquameter
®
The AP-Lite is designed to connect to the Aquameter
®
using an EX-2000 Aquaprobe
®
Extension Cable. The Aquaprobe
®
Extension Cable features high-pressure metal
connectors, which incorporate several O-ring seals at the Probe end. Prior to first
connection, the seals must be lubricated using the silicone grease provided.
Apply a generous smear of grease to the O-rings where indicated above. Be careful not to
get any grease inside the connector near the gold contacts. A small smear of grease
should also be applied to the thread on the Probe to allow easy tightening of the collar.
To connect the Extension Cable to the AP-Lite, align the white dot on the AP-Lite with the
◄AQUAREAD
logo on the plug body, then press the plug into the socket and tighten the
retaining collar fully.
DO NOT TWIST THE CONNECTOR BODY WITH RESPECT TO THE
PROBE
. Once the AP-Lite has been connected to the Extension Cable, the Aquameter
®
can be connected.
Always ensure the Aquameter
®
is switched off prior to connecting or disconnecting an AP-
Lite. Align the
◄AQUAREAD
logo on the plug body with the red on/off switch on the
Aquameter
®
, then press the plug into the socket and tighten the retaining collar.
Once the AP-Lite is connected to the Aquameter
®
, switch the Meter on by pressing the red
on/off switch.
The Aquameter
®
should detect the Probe and automatically start displaying readings.