![Aquaread Aquaprobe Lite Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/aquaread/aquaprobe-lite/aquaprobe-lite_instruction-manual_2960646039.webp)
Aquaprobe
®
Lite Instruction Manual
10401-00976 Rev F
Press the
OK
key to continue. The reading on the Aquameter
®
directly after calibration
should be approximately 118 µg/L at 20ºC (this value will vary with temperature).
Calibration is now complete.
11.6.8.
Errors During Calibration
If a problem occurs during calibration, an error message will be displayed. Refer to
in section 10 for error message handling.
11.6.9.
Lens and Sleeve Maintenance
On a daily basis, the lenses on the electrode should be wiped over with a soft damp cloth.
Similarly, the inside of the Probe Sleeve and Sleeve Cap should be kept clean and free
from any deposits that may cause stray fluorescence.
Never use an abrasive cleaner on the inside of the Probe sleeve or cap as they have
been treated with a non-reflective coating which can be easily damaged.
The inside of
the sleeve should be wiped over with a soft damp cloth and
non-abrasive
detergent.
Always re-calibrate the zero point after cleaning the sleeve or lenses.
© 2015 Aquaread
®
Ltd.
www.aquaread.com
Page 39 of 72