33
CAUTION
, Totalizer value is lost.
[E-EEPTRP]
EEPROM error in Trip Totalizer. Trip Totalizer value may be incorrect.
Action: Go to [DEFAULTS] correct Trip Totalizer memory block (under USER Code).
CAUTION,
Trip Totalizer is lost.
[E-EEPFAT]
EEPROM error in File System.
Action: electronic module must be replaced. Please contact the supplier.
7.2
Alarm messages VZF II / VZFA II
The electronic module performs a self-test about every 5 minutes. If an alarm condition is
detected,
[ALARM]
message will appear on the display every 2 seconds.
Alarm messages are messages from the parameter settings
[nO ALmS]
no alarm is active
Action: none.
[A-SEnSSI]
alarm when sensor simulation is activated (Service Technician only).
Action: none. CAUTION Totalizer and Trip Totalizer accumulate the simulated value.
[A-o1 SI]
alarm when output 1 simulation is activated.
Action: turn off when not required anymore.
[A-o2 SI]
alarm when output 2 simulation is activated.
Action: turn off when not required anymore.
[A-ANA SI]
alarm when Analog current loop outputsimulation is activated.
Action: turn off when not required anymore.
[A-o1 LI]
alarm when output 1 is over its limit.
Action: adjust the settings of output1.
[A-o2 LI]
alarm when output 2 is over its limit.
Action: adust the settings of output2.
[A-AnA LI]
alarm when Analog current loop outputis over its limit.
Action: adust the settings of analog current loop output.