Appendix A
Moving the 7300/7500/7500 Fast System
116
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
5.
Run a TaqMan
®
RNase P Instrument Verification
plate (see
).
• If the run passes, recalibrations are not
necessary.
• If the run fails, perform steps
to recalibrate the instrument.
6.
Perform an ROI calibration (see
).
7.
Perform a Background calibration (see
).
Perform an optical calibration if you are using a
7500 or 7500 Fast system (see
8.
Perform a pure dye calibration (see
9.
Perform an instrument verification run (see