CSPP Capacitor Charging Unit Operation Manual
CSP-CSPP-8000/2
_________________________________________________________________________________________________________
__________________________________________________________________________________________________________
Page
10
of
59
Output Cable Connection Cont
The junction box is usually supplied with a 1.5 metre lead, terminated in our HV connector. The
junction box is fitted with a microswitch which is connected to sense pins in the HV plug, ie
deactivating the high voltage charger if the lid is removed and th
e „INTERLOCK‟ light will
illuminate.
Line Voltage Connection
Mains input is connected by the 3-Pin Amphenol on the left of the front panel. Nominal input
voltage is 230VAC 45-65 Hz. The wiring to this connector is as follows:-
A
Ground / Earth
B
Neutral
C
Live
If a ready-wired plug is supplied, it will be supplied from the factory with the following colour
coding:-
Green/Yellow
Earth
Blue
Neutral
Brown
Live
The operator must ensure that the ac supply is capable of supplying sufficient energy to power the
CSP-P unit. Although the unit will operate from most generators of 3 kVA, the quality and
regulation vary considerably from make to make and how well they have been maintained. We
have used 3.5 kVA generators without problems. A voltage stabiliser is not necessary as any
reasonable variation of the AC supply will not affect the output voltage (hence power) unlike older
designs.
NB Some „inverter‟ type generators do not work well with CSP-P units.