(Advanced Settings)
AUTO
(Auto Off): The Auto-Off feature will automatically power off the MND
whenever no buttons are pressed within the specified time frame. The minimum Auto-
Off time is 15 seconds. To disable the Auto-Off feature, set the value to 65535 (default).
NOTE: the Auto-Off cannot be set to less than the relay “On Time” setting when using
the Timed Relay option (see Outputs).
dEC PL
(Decimal Place): defines where the decimal point will be displayed within
the reading.
NOTE: most APG sensors have the ability to set the number of decimal places of the
readings being sent to the MND. Refer to the sensor manual for more information.
MASK
(Digit Mask): allows the user to mask the value of the least significant digit(s),
up to 3 places, so that masked digit(s) will always display 0 and will not increment.
SHIFT
(Digit Shift): allows the user to shift the reading to the right by dividing the
reading by 10, 100, or 1000.
MLTPLR
(Multiplier): allows the user to apply a conversion multiplier to the sensor
readings. For example, a multiplier setting of 2.0 will double the reading received from
the sensor.
NOTE: most APG sensors have the capability to apply a conversion multiplier to the
reading before the value it is sent to the MND.
bAR 0
(Bar Graph 0%) &
bAR100
(Bar Graph 100%): define the readings
associated with 0% and 100% on the display bar graph. Bars will appear/disappear in
10% increments of the total span.
NOTE: the bar graph limits will be applied to all sensor readings when multiple sensor
are being displayed, and cannot be set individually for each sensor.