10. ORP Measurement 

ORP  stands  for  Oxidation-Reduction  Potential,  measured  in  mV.  It’s  also  called  redox.  ORP is  a 
measure of the cleanliness of water & its ability to break down contaminants. A separate ORP probe 
(ORP60-DA) needs to be installed to be able to measure ORP.   
 
Power on the PC60-Z tester, unscrew the original probe, and install the ORP60-DA probe, then the 
tester will automatically switch to ORP measurement mode (Refer to Section 14 for how to replace 
a probe). 
 
Rinse the probe in distilled water and dry it. Dip the probe in sample solution, shake for a few seconds, 

and allow it to stand still. Get the ORP readings after 

appears and stays on screen. 

 

11. Probe Cleaning 

1)  The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before 

and after each measurement with pure water in a container or with a wash bottle.   

2)  For  tough  contaminants,  detach  the  sensor  shield,  soak  the  probe  in Apera  probe  cleaning 

solution  (AI1166)  or  detergent  water  for  30  minutes.  Then  use  a  soft  brush  to  remove  the 
contaminants. Afterwards, soak the probe in 3M KCL soaking solution for at least 1 hour. Rinse it 
off, then re-calibrate the tester before using again. 

 

12. Probe Storage 

1)  Under regular usage (daily or weekly use), make sure the probe cap is wet, and tightly close the 

cap with the O-ring. 

2)  For long-term storage (you are not going to use the product for a while), add 3M KCL soaking 

solution to the Fill line in the probe cap and store the probe in it. Close on the probe cap tightly 
with the O-ring. 

3)  If you find white crystals inside or outside the probe cap, it is perfectly normal. It is the 3M KCL 

soaking solution that crystalizes over time by its nature. Just rinse them off and add in new soaking 
solution. This chemical is not poisonous nor dangerous, and the probe’s performance will not be 
affected at all. 

4)  NEVER  store  the  probe in pure  water  like tap,  RO,  distilled,  or  deionized  water  as they could 

damage the pH probe. If this happens, immediately soak the pH probe Apera 3M KCL soaking 
solution overnight, then re-calibrate it before using. Pure water is only for rinsing the probe. 

Содержание AI3719

Страница 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Страница 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Страница 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Страница 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Страница 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Страница 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Страница 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Страница 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Страница 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Страница 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Страница 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Страница 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Страница 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Страница 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Страница 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Отзывы: