13. Parameter Setting 

 Table of Settings 

 

 Parameter Setting 

1) 

When the meter is turned off, long press 

  to enter parameter setting →short press 

  to 

switch  P01-P02…  →P14.  Short  Press 

 ,  parameter  flashes  →  short  press 

   to  adjust 

parameter  →  short  press 

  to  confirm →Short  press 

  to  exit  parameter setting  and  go 

back to measurement mode. 

2) 

Auto. Lock (P02)

 – Users can set the auto lock time from 5 to 20 seconds. For example, if 10 

seconds is set, when the measured value is stable for more than 10 seconds, the measured value 

will be automatically locked, and the HOLD icon will be displayed. Short press 

to release the 

lock. When the setting is "Off", the Auto. lock function is turned off, that is, the measured value 

can only be manually locked. Short press 

to lock or unlock the measured value. The HOLD 

icon will be displayed when reading is locked. 

3) 

Auto.  Backlight  (P03)

 

  Users  can  set  the  automatic  backlight  time  for  1  to  8  minutes.  For 

example, if 3 minutes is set, the backlight will turn off automatically after 3 minutes; when the “Off” 

is  set, the  auto.  backlight  function  will  be  turned off,  and  short  press 

to  manually  turn the 

backlight on or off. 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C – ˚F 

˚F 

P2 

Select automatic lock 

5-20 seconds 

– Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

– Off 

P4 

Automatic Power Off 

10-20 minutes 

– Off 

10 

P5 

pH Buffer Series Selection 

USA 

– NIST 

USA 

P6 

pH Resolution 

0.1 

– 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in 

ZenTest App) 

P8 

pH back to factory default 

No

 

– Yes 

No 

P9 

Conductivity Reference 

Temperature 

1

5 ˚C to 30 ˚C 

2

5 ˚C 

P10 

Temp. Compensation Coefficient 

0 to 9.99 

2.00 

P11 

Conductivity Calibration Reminder 

H-hours D-Days (set up in 

ZenTest App) 

P12 

Conductivity Back to Factory 

Default 

No 

– Yes 

No 

P13 

TDS Factor 

0.40 to 1.00 

0.71 

P14 

Salinity Unit 

ppt 

– g/L 

ppt 

Содержание AI3719

Страница 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Страница 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Страница 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Страница 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Страница 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Страница 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Страница 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Страница 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Страница 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Страница 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Страница 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Страница 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Страница 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Страница 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Страница 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Отзывы: