background image

 

- 5 - 

4.  Keypad Functions 

 

 

Short

 

press------

 < 2 seconds ,

Long press-------

 > 2 seconds

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

5.  Preparation before Use 

  Pull out the battery insulation slip, and take off the probe cap.

 

  Rinse off the probe in pure water (preferably distilled or deionized water. RO water or tap 

water is the alternative) , then shake off excess water. 

 
 
 
 
 
 
 
 
 
 
 
 

  Perform calibration. For pH calibration tutorial, refer to Section 6; for conductivity calibration, 

refer to Section 8.

 

 

If the tester hasn’t been used for a long time (over 1 month), please soak the probe in the 3M 
KCL soaking solution for 15 minutes, then calibrate it before test.

 

 

 

1. When turned off, short press to turn on the tester; long 

press to enter parameter setting. 

2. In calibration mode or parameter setting, short press to 

return to measurement mode. 

3. In measurement mode, long press to turn off the tester, 

short press to turn on/off backlight. 

 

1.In measurement mode, short  press to  switch  parameter 

pH

Cond

TDS

Sal

Res 

2.In  measurement  mode,  long  press  to  turn  on/off 

Bluetooth

®

 receiver. When turned on, 

  will be flashing; 

when connected to smartphone,   will stay on. 

3.In  parameter  setting,  short  press  to  change  parameter 

(Uni-directional). 

 

1. Long press to enter calibration mode. 
2. In calibration mode, short press to confirm calibration. 
3. In measurement mode, when automatic lock is turned off, 

short press to manually lock or unlock readings. 

Battery Cap

 

Buttons

 

Probe 

 

PC60-DE 
Probe 

Probe Cap

 

LCD Display

 

pH Sensor 

Cond/TDS/Sal/Res Sensor 

Probe Ring 

 

Содержание AI3719

Страница 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Страница 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Страница 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Страница 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Страница 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Страница 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Страница 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Страница 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Страница 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Страница 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Страница 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Страница 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Страница 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Страница 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Страница 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Отзывы: