15. Icons and Functions 

 
 
 
 
 
 
 
 
 
 

16. Probe Replacement 

To replace a probe:   

1) Take off the probe cap; 2) Screw off the probe ring 3) Unplug the probe; 4) 
Plug in the new replacement probe (pay attention to the probe’s position); 5) 
Screw on the probe ring tightly. Soak the probe in 3M KCL for 5-15 minutes. 
Then perform calibration before testing.

 

 
The replacement probes that are compatible with PC60-Z are: 

 

PC60-DE (Default pH/cond. Probe),   

 

PH60-DE (Regular pH glass bulb probe), PH60S-DE (Spear pH 
probe for solids/semi-solids pH testing), PH60F-DE (Flat pH probe for surface pH testing),   

 

EC60-DE (Conductivity probe) 

 

ORP60-DA (ORP Probe). 

 
*Probes don’t last forever. Every probe will eventually age and fail even if you don't use it that often. 
The typical service life of Apera probes is 12-24 months depending on the frequency of usage and 
how well you keep it clean and properly stored. We recommend replacing your probe every 1 to 2 
years to ensure the best performance. 
 

17. Troubleshooting Guide 

Calibrated points 

 

Self-Diagnosis 

Symbol 

Er1, Er2, Er3, Er4,Er5, Er6 

Stable reading 

indicator 

 

Waterproof Rating 

IP67, floats on water 

Reading Lock 

HOLD 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal 

 

Battery Life 

200 Hours 

Low power reminder 

           

Backlight 

White: Measurement;   

Green: Calibration; Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/680g; 

Graph-3 

LCD Display 

Graph-4 

pH calibration reminder 

 

Graph-5 

pH alarm triggered 

 

Содержание AI3719

Страница 1: ...1 SKU AI3719 ZenTest PC60 Z Smart 7 in 1 Tester pH Cond TDS Salinity Resistivity ORP Temp Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...be Replacement 14 17 Troubleshooting Guide 14 18 Warranty 16 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attrib...

Страница 3: ...er to the functions available on each platform in the following table This manual shows you how to operate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mob...

Страница 4: ...an be added to saved data Data Output N A Share data via Email 2 What s in the Kit 3 Battery Installation Please install batteries according to the following steps Please note direction of batteries A...

Страница 5: ...tester long press to enter parameter setting 2 In calibration mode or parameter setting short press to return to measurement mode 3 In measurement mode long press to turn off the tester short press to...

Страница 6: ...oint calibrations 4 00 10 01 1 68 or 12 45 immediately after the 1st point calibration is finished Do NOT turn off the meter before you calibrate 2nd or 3rd point Otherwise you will need to restart th...

Страница 7: ...pH electrode is damaged or broken If so please replace with a new one 2 The glass bulb or junction is severely contaminated Please use a soft brush with soap water to clean it thoroughly Then soak it...

Страница 8: ...ur certain amount of 1413 S cm and 12 88mS cm conductivity calibration solution into corresponding calibration bottles to about half volume of the bottle 3 Long press key to enter calibration mode scr...

Страница 9: ...tion is expired or contaminated 3 Check if the conductivity electrode two black rods is damaged 4 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to c...

Страница 10: ...ductivity is linear related and its conversion factor is 0 40 to 1 00 Adjust the factor in parameter setting P5 based on the requirements in different industries The factory default setting is 0 71 d...

Страница 11: ...I1166 or detergent water for 30 minutes Then use a soft brush to remove the contaminants Afterwards soak the probe in 3M KCL soaking solution for at least 1 hour Rinse it off then re calibrate the tes...

Страница 12: ...ding is locked 3 Auto Backlight P03 Users can set the automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off...

Страница 13: ...alibration is finished or the reminder setting is cancelled in the ZenTest App the Er6 icon will disappear 6 pH Back to Factory Default P08 and Conductivity Back to Factory Default P12 Select Yes to r...

Страница 14: ...ry probe will eventually age and fail even if you don t use it that often The typical service life of Apera probes is 12 24 months depending on the frequency of usage and how well you keep it clean an...

Страница 15: ...tabilizing add 1ml of 3M KCL solution to 1000ml of test solution or 1 teaspoon to 1 gallon Display similar readings in any solutions or always display 7 0 pH Broken probe If you don t find any visible...

Страница 16: ...PERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized repair normal w...

Отзывы: