4    Complete Kit 

 

 

 

 

 

 

 

 

5  Things to Know Before Use 

1) 

Be careful about the glass tip of the probe. It can break if being contacted by external force. If 
broken, the probe needs to be replaced.

 

2) 

Things needed in addition to what’s in the box:

 

a) 

Distilled or deionized water (8-16oz) for rinsing the probe after each test

 

b) 

Tissue paper for drying the probe

 

3) 

For PH60-Z and PH60F-Z testers:

 

a) 

A few drops of water are added to the probe cap to keep the pH electrode in an activated 
state before the tester leaves factory. Generally, users can start using the tester directly. If 
the measuring response is slow, users can soak the pH electrode for 30 minutes in the 3M 
KCL storage solution (fill to the fill-line) before using it; if the electrode is dry for a long 
time (> 1 month), the pH electrode responds slowly, You can soak the pH electrode for 8 
hours in the storage solution before using.

 

b) 

When the tester is not in use, we recommend adding one to two drops of tap water to the 
probe  cap,  and  close  the  cap  (be  careful  not  to  add too  much  water). This  way,  the  pH 
probe’s sensitivity can be maintained and users can start using the tester right away next 
time.

 

c) 

Do not soak the pH electrode in purified (e.g. distilled/deionized) water for a long time, which 
will make the electrode response slow. If this happens, soak the pH electrode in 3M KCl 
solution for 3~5 hours, and then re-calibrate it before using.

 

d) 

The storage solution is 3M KCL (SKU: AI1107), and the tester kit comes with a bottle of 
10mL storage solution (can be used repeatedly). If it is contaminated, replace it with a new 
one. Please do not use other brands’ storage solutions as they may contain other chemicals 
that can cause damage to the electrodes.

 

4) 

For PH60S-Z Spear pH Tester:

 

 

NEVER

 store the spear probe in a dry condition  because  permanent  damage 

CAN

  be 

caused. The spear probe should always be stored in the 3M KCL soaking solution.

 

Содержание 60-Z Series

Страница 1: ...1 60 Z Series Smart pH Tester pH Temp ORP Instruction Manual PH60 Z pH Tester PH60S Z Spear pH Tester PH60F Z Flat pH Tester www aperainst de V1 0...

Страница 2: ...Probe Replacement 11 13 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leave...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone followin...

Страница 4: ...uetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode...

Страница 5: ...the storage solution before using b When the tester is not in use we recommend adding one to two drops of tap water to the probe cap and close the cap be careful not to add too much water This way th...

Страница 6: ...y rinse the probe in distilled water and dry it and place the probe in 10 01 buffer solution sold separately to complete 3rd point of calibration according to the steps in 6 will appear at the bottom...

Страница 7: ...pressed before measurement is fully stable Wait for to come up and stay on screen before pressing During calibration readings being unstable for over 3 minutes 1 Please check whether pH electrode is...

Страница 8: ...lid medium When performing such tests pay attention to insert probe evenly and be careful to prevent probe from damage If the medium is too hard such as meat or fruits please bore a small hole with a...

Страница 9: ...screw the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14 for how to replace a probe Rinse the probe in distilled wa...

Страница 10: ...on will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Reminder On the meter y...

Страница 11: ...onsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized...

Отзывы: