- 11 -

 

11 Icons and Functions 

 

12 Probe Replacement 

To replace a probe:   

1)  take off the probe cap; screw off the probe ring; unplug the probe 
2) 

plug in the new replacement probe (pay attention to the probe’s position)

 

3)  screw on the probe ring tightly.   

The replacement probes that are compatible with PH60F-Z are: 

  PH60-DE (Regular pH glass bulb probe),   

  PH60S-DE (Spear pH probe for solids/semi-solids pH testing),   

  PH60F-DE (Flat pH probe for surface pH testing) 

  ORP60-DA (ORP Probe). 

13 Warranty 

We warrant this instrument to be free from defects in material and workmanship and agree to repair 
or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned 
or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a 
period of TWO YEARS (SIX MONTHS for the probe) from the delivery. This limited warranty does 
NOT cover any damages due to: accidental damage, unauthorized repair, normal wear and tear, or 
external causes such as accidents, abuse, or other actions or events beyond our reasonable control.

 

 

 
 
 
 
 
 
 

APERA INSTRUMENTS (Europe) GmbH

 

Wilhelm-Muthmann-Straße 18, 42329 Wuppertal, Germany

 

Contact: 

[email protected] 

www.aperainst.de

 | Tel. +49 202 51988998

 

Calibrated points

 

 

Self-Diagnosis 

Symbol

 

Er1, Er2, Er3, Er4,Er5, Er6

 

Stable reading 

indicator

 

 

Waterproof 

Rating 

IP67, floats on water 

Reading Lock

 

HOLD

 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal

 

 

Battery Life 

200 Hours 

Low power 

reminder

 

           

Backlight 

White: Measurement

Green: 

Calibration

Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/680g; 

Содержание 60-Z Series

Страница 1: ...1 60 Z Series Smart pH Tester pH Temp ORP Instruction Manual PH60 Z pH Tester PH60S Z Spear pH Tester PH60F Z Flat pH Tester www aperainst de V1 0...

Страница 2: ...Probe Replacement 11 13 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leave...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone followin...

Страница 4: ...uetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode...

Страница 5: ...the storage solution before using b When the tester is not in use we recommend adding one to two drops of tap water to the probe cap and close the cap be careful not to add too much water This way th...

Страница 6: ...y rinse the probe in distilled water and dry it and place the probe in 10 01 buffer solution sold separately to complete 3rd point of calibration according to the steps in 6 will appear at the bottom...

Страница 7: ...pressed before measurement is fully stable Wait for to come up and stay on screen before pressing During calibration readings being unstable for over 3 minutes 1 Please check whether pH electrode is...

Страница 8: ...lid medium When performing such tests pay attention to insert probe evenly and be careful to prevent probe from damage If the medium is too hard such as meat or fruits please bore a small hole with a...

Страница 9: ...screw the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14 for how to replace a probe Rinse the probe in distilled wa...

Страница 10: ...on will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Reminder On the meter y...

Страница 11: ...onsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized...

Отзывы: