3)  PH60F-Z Flat probe testers are mostly for flat surface sample test. 

a) 

For skin test: skin should be without sweat or dirt, nor be overly cleaned (do not use facewash 
products before testing) to avoid affecting measurement results, dampen skin with some distilled 
water, slightly force flat probe onto the skin, get readings after value stabilized.

 

b) 

For paper, fabric and leather test: add a few drops of distilled/deionized water on surface before 
performing tests.

 

c) 

For micro sample testing, use a container with an inner diameter<=19mm and a flat bottom. The 
tester can test volume >=0.5ml.

 

8  ORP Measurement 

ORP stands for Oxidation-

Reduction Potential, measured in mV. It’s also called redox. ORP

 is a 

measure of the cleanliness of water & its ability to break down contaminants. A separate ORP 
probe (ORP60-DA) needs to be installed to be able to measure ORP.   
 
Power on the PH60-Z tester, unscrew the original probe, and install the ORP60-DA probe, then the 
tester will automatically switch to ORP measurement mode (Refer to Section 14 for how to replace 
a probe). 
 
Rinse the probe in distilled water and dry it. Dip the probe in sample solution, shake for a few 
seconds, and allow it to stand still. Get the ORP readings after 

  appears and stays on screen. 

9  Parameter Setting 

9.1  Table of Settings 

9.2  Parameter Setting 

1)  When the meter is turned off, long press 

 

to enter parameter setting →short press 

  to 

switch P01-

P02… →P14. Short Press 

, parameter flashes → short press

 

  to adjust 

parameter → short press 

 

to confirm →Short press 

  to exit parameter setting and go 

back to measurement mode. 

Symbol 

Parameter Setting Contents 

Content 

Factory Default 

P1 

Temperature Unit 

˚C –

 

˚F

 

˚F

 

P2 

Select automatic lock 

5-20 seconds 

 Off 

Off 

P3 

Automatic Backlight Off 

1-8 minutes

 

 Off 

P4 

Automatic Power Off 

10-20 minutes 

 Off 

10 

P5 

pH Buffer Series Selection 

USA 

 NIST 

USA 

P6 

pH Resolution 

0.1 

 0.01 

0.01 

P7 

pH Calibration Reminder 

H-hours D-Days (set up in 

ZenTest App) 

P8 

pH back to factory default 

No

 

 Yes 

No 

Содержание 60-Z Series

Страница 1: ...1 60 Z Series Smart pH Tester pH Temp ORP Instruction Manual PH60 Z pH Tester PH60S Z Spear pH Tester PH60F Z Flat pH Tester www aperainst de V1 0...

Страница 2: ...Probe Replacement 11 13 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leave...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone followin...

Страница 4: ...uetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode...

Страница 5: ...the storage solution before using b When the tester is not in use we recommend adding one to two drops of tap water to the probe cap and close the cap be careful not to add too much water This way th...

Страница 6: ...y rinse the probe in distilled water and dry it and place the probe in 10 01 buffer solution sold separately to complete 3rd point of calibration according to the steps in 6 will appear at the bottom...

Страница 7: ...pressed before measurement is fully stable Wait for to come up and stay on screen before pressing During calibration readings being unstable for over 3 minutes 1 Please check whether pH electrode is...

Страница 8: ...lid medium When performing such tests pay attention to insert probe evenly and be careful to prevent probe from damage If the medium is too hard such as meat or fruits please bore a small hole with a...

Страница 9: ...screw the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14 for how to replace a probe Rinse the probe in distilled wa...

Страница 10: ...on will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Reminder On the meter y...

Страница 11: ...onsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized...

Отзывы: