Apera Instruments 60-Z Series Скачать руководство пользователя страница 10

 

- 10 -

 

2) 

Auto. Lock (P02)

 

 Users can set the auto lock time from 5 to 20 seconds. For example, if 10 

seconds is set, when the measured value is stable for more than 10 seconds, the measured 
value will be automatically locked, and the HOLD icon will be displayed. Short press 

  to 

release the lock. When the setting is "Off" , the Auto. lock function is turned off, that is, the 
measured value can only be manually locked. Short press 

  to lock or unlock the measured 

value. The HOLD icon will be displayed when reading is locked. 

3) 

Auto. Backlight (P03)

 

  Users can set the automatic backlight time for 1 to 8 minutes. For 

example, if 3 minutes is set, the backlight will turn off automatically after 3 minutes; when the 

Off

 is set, the auto. backlight function will be turned off, and short press 

  to manually 

turn the backlight on or off. 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For 

example, if 15 minutes is set, the meter will automatically shut down after 15 minutes if no 
operation; when 

Off

” i

s set, the auto. power off function will be turned off. Long press 

  to 

manually shut down the meter. 

5) 

pH Calibration Reminder (P07) 

 set X hours (H) Or X days (D) in ZenTest mobile app 

 

settings 

 Parameter 

 pH 

 Calibration Reminder. On the meter, you can only check the 

values that’s been set up on ZenTest App. For example, if 3 days is set up, the Er6 icon (see 

Figure-4) will appear in the lower right corner of the LCD screen in 3 days to remind you to 
perform calibration, also in the ZenTest App there will be a pop-up reminder. After calibration 
is finished or the reminder setting is cancelled in the ZenTest App, the Er6 icon will disappear. 

6) 

pH Back to Factory Default (P08) 

 

Select “Yes” to recover instrument calibration to 

theoretical value. This function can be used when instrument does not work well in calibration 
or measurement. Calibrate and measure again after setting the instrument back to factory 
default.   

10 Technical Specifications 

pH 

Range 

-2.00 to 16.00 pH 

Resolution 

0.01 pH 

Accuracy 

±0.01 pH ±1 digit 

Calibration Points 

1 to 3 points 

Auto. Temperature Compensation 

 

50˚C (32 –

 

122˚F)

 

ORP (mV) 

Range 

-1000 mV to 1000 mV 

Accuracy 

±0.2% F.S 

Temperature 

Range 

0 to 50˚C (32

-

122˚F)

 

Accuracy 

±0.5˚C

 

 

 

 

 

 

 

 

Graph-3

 

LCD Display

 

Graph-4

 

pH calibration reminder

 

 

Graph-5

 

pH alarm triggered

 

 

Содержание 60-Z Series

Страница 1: ...1 60 Z Series Smart pH Tester pH Temp ORP Instruction Manual PH60 Z pH Tester PH60S Z Spear pH Tester PH60F Z Flat pH Tester www aperainst de V1 0...

Страница 2: ...Probe Replacement 11 13 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the pH sensor before the product leave...

Страница 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone followin...

Страница 4: ...uetooth receiver When turned on will be flashing when connected to smartphone will stay on 2 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode...

Страница 5: ...the storage solution before using b When the tester is not in use we recommend adding one to two drops of tap water to the probe cap and close the cap be careful not to add too much water This way th...

Страница 6: ...y rinse the probe in distilled water and dry it and place the probe in 10 01 buffer solution sold separately to complete 3rd point of calibration according to the steps in 6 will appear at the bottom...

Страница 7: ...pressed before measurement is fully stable Wait for to come up and stay on screen before pressing During calibration readings being unstable for over 3 minutes 1 Please check whether pH electrode is...

Страница 8: ...lid medium When performing such tests pay attention to insert probe evenly and be careful to prevent probe from damage If the medium is too hard such as meat or fruits please bore a small hole with a...

Страница 9: ...screw the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14 for how to replace a probe Rinse the probe in distilled wa...

Страница 10: ...on will be turned off Long press to manually shut down the meter 5 pH Calibration Reminder P07 set X hours H Or X days D in ZenTest mobile app settings Parameter pH Calibration Reminder On the meter y...

Страница 11: ...onsibility of APERA INSTRUMENTS Europe GmbH for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to accidental damage unauthorized...

Отзывы: