APAR AR911 Скачать руководство пользователя страница 1

 
 

 

 

 

User instruction

 

AR911

 

 

 

 

 

 

 

 

 

 

 

Setter - analog signal 

measuring device

 

 

 

 

Version 1.0.2 

2013-03-20

APAR - SALES OFFICE

 

05-090 Raszyn, ul. Gałczyńskiego 6 
Tel.: +48 22 101-27-31, +48 22 853-49-30 
E-mail: 

[email protected] 

Website:    

www.apar.pl

 

 

Содержание AR911

Страница 1: ...ruction AR911 Setter analog signal measuring device Version 1 0 2 2013 03 20 APAR SALES OFFICE 05 090 Raszyn ul Ga czy skiego 6 Tel 48 22 101 27 31 48 22 853 49 30 E mail automatyka apar pl Website ww...

Страница 2: ...ER 3 3 CONTENTS OF THE SET 4 4 TECHNICAL DATA 4 5 DIMENSIONS OF THE ENCLOSUREAND DESCRIPTION OF CONNECTIONSAND ELEMENTS 5 6 BUTTON FUNCTIONS 6 7 RANGE OF INDICATIONS CHANGE OFAVALUE SET FOR AN OUTPUT...

Страница 3: ...measuring standard current and voltage signals analog output input setter or measuring device current 0 4 20 mA active output may not be supplied in a two wire current loop voltage 0 2 10 V the outpu...

Страница 4: ...mperature changes 0 005 of the input range C Output input response time 10 90 0 36 s output 0 32 1 3 s input programmable with parameter 6 FiLt A 7 segment LCD display without background illumination...

Страница 5: ...TION OF CONNECTIONS AND EXTERNAL ELEMENTS 1 Enclosure Enclosure type manual portable MBMO series Material ABS Enclosure dimensions 162 x 89 x 25 mm Protection rating IP43 IP20 on the connection side 2...

Страница 6: ...chapter 8 SET UP a stepwise limit change of the set value of the output signal upper value of the indication range 20 mA 10 V parameter 4 rtoP or setting range reduction 8 LiLo SET DOWN a stepwise li...

Страница 7: ...THE CONFIGURATION PARAMETERS All the configuration parameters of the device are stored in the non volatile EEPROM internal memory The parameters are set from the film keypad located on the front panel...

Страница 8: ...tput setting mode also set using the OUT button 8 LiLo lower setting limit 999 9999 units lower limit of the set value parameter 11 SEt when setting using the setter buttons 9 99 9 LiHi upper setting...

Страница 9: ...es are different than zero 12 riSE and 13 FALL a periodic triangular signal is generated on the outputs The limit values amplitudes of the output signal are equal to the limit values of the signal use...

Страница 10: ...errors one time and recurring Code Description of message the operating mode was initiated input output PoFF the calibrator was switched off manually or automatically due to low battery voltage CodE...

Страница 11: ...11 11 USER S NOTES...

Страница 12: ...12...

Отзывы: