Product Operation Technical Tips
Monitor terminal display information
•
There will inevitably be more or less positive and negative ions in the detection space which will also have a certain impact on the detection results
due to the existence of cosmic rays, trace radioactive substances in the environment, and the use of various electrical equipment.
Working status: monitor whether the sensor is working normally and alarm output
Equipment Address: Display the address set by the sensor
Detection distance: Display the detection distance set by the electrostatic sensor
Threshold voltage: Displays the set static voltage safety (alarm) threshold
Real-time voltage: display the static voltage value on the surface of the measured object
IMS Smart Controller
P3
P2
WAN/LAN P1
P4 POWER
P1-P4 : 1/2-VCC 3-RS485_B 4-RS485_A 5/6-GND 7-C_V
TH
/HV
AL
8-C_ACT/Clean 9-PE(Metal shield)
P1
P2
P4
P3
Red-V
TH
/HV
AL
ACT/Clean-Green
ACT
Work/Link
NEXT
SCAN
NEXT
LINK
B
SET
S
SCROLL
T
RJ45 port
2.4G communication antenna
Wireless connection
WAN/LAN
:
Ethernet connection port
M4 mounting holes
Temperature and humidity detection window
•
Aim at the LED on the front of the product during operation (the distance should not exceed 1m), first press the unlock key, and then press the
corresponding function key to set, the red light flashes when the key is pressed.
•
The calibration plate should be much larger than the sensor detection window, and the plate and sensor should be well grounded when performing
zero adjustment.
•
The calibration plate should be much larger than the sensor detection window, and the sensor should be well grounded when performing calibration.
•
There should be no obstruction between the sensor and the detected object, otherwise the accuracy of the detection result will be affected.
•
There should be no electrical equipment that may affect the sensor within the detection distance of the sensor.
•
The plane of the sensor detection window must be parallel to the surface of the object to be measured to accurately measure charged objects.
•
The measured value is smaller than the actual electrostatic value of the charged object when the charged object is smaller than the calibration plate.
•
The measured value is larger than the actual electrostatic value of the charged object when the charged object is larger than the calibration plate.
•
Do not adjust zero in the static charge state or in the electrostatic detection process (Non-electrostatic calibration process); the displayed static value
will be zero if reset it during static electricity testing.
•
Influence of temperature and humidity on electrostatic detection:
①
The lower the temperature, the lower the humidity, the less moisture in the space, the easier the surrounding objects are to be electrified by
friction, and the greater the interference to electrostatic detection.
②
The higher the temperature, the higher the humidity, the more moisture in the space, the more active the movement of water molecules, the more
easily the calibration device generates corona or spark discharges, and the greater the impact on the uniform electric field generated by the
calibration device which will weaken the uniform electric field.
③
The lower the temperature, the less moisture in the space, the easier the surrounding objects are to be electrified by friction, and the greater the
impact on electrostatic detection under the same humidity.
Therefore, the ambient temperature and humidity during calibration/testing should be clearly marked when doing electrostatic calibration/testing.
LCD
Electrostatic sensor monitoring settings
Temperature and humidity display
Wrok/Link
:
Network connection
:
red light flashing, green light always on
Network disconnection
:
red light flashing, green light is off
Electrostatic eliminator running time setting
Electrostatic eliminator monitoring settings
ACT
:
Monitor power-on indication, green light always on
P1→P4
:
4 ways status indicator light
Red light flashing: Sensor threshold alarm/eliminator high voltage alarm
Green light always on: Sensor/eliminator power-on indication
Green light flashing: Eliminator cleaning indication
POWER
:
Monitor power port
(with RS485 communication)
P1→P4
:
The working status monitoring and
power supply port of the eliminator or sensor