DIGITAL CALIBRATION
SOLDER IRON DIGITAL TEMP CALIBRATION
By default, the system is properly calibrated but for cases when a little
adjustment of the soldering iron calibration is required the following
procedure can be done.
1. Turn on the soldering iron function switch.
2. Set to appropriate temperature you want to calibrate. Place the tip of
the soldering iron on an external temperature sensor.
3. The readings on the external temperature sensor should be more or
less equal to the displayed temperature.
4. If there are large discrepancy in the temperature reading we can re-
calibrate the temperature setting.
5. While the solder iron is operating make sure the hot air gun and
Preheating function is in off mode (“
0FF
” is displayed on the panels
“1” and “2”), hold the preheater UP button for a few seconds until
four zeroes are displayed “0000” .
6. Adjust the temperature compensation using the UP and DOWN
buttons of the Soldering iron adjustment buttons.
7. A zero “0” on the first digit signifies addition to the current
temperature and a minus “-” on the first digit will subtract the
displayed value from the current settings.
8. Confirm the change by pressing the air pressure down button.