- 16 -
1
√
2
3
4
F1.7
(
Digital Filter 1
)
5
0
1
2
√
3
4
F1.8
(
Digital Filter 2
)
5
0
1
2
√
3
4
F1.9
(
Digital Filter 3
)
5
higher the value, the lower is the effect acting upon
the indicator due to noise and mechanical vibration
and hence having a more accurate display.
However, it slows down the settling rate of the
indicator.
0
(
2OUT
)
1
(
4OUT
)
√
2
(
8OUT
)
3
(
16OUT
)
4
(
32OUT
)
5
(
64OUT
)
F1.10
(
Digital filter
cutout sensitivity
)
6
(
128OUT
)
Digital Filter Cutout Sensitivity. Specifies the
number of consecutive readings that must fall
outside the filter threshold (the value set for F1.11)
before digital filtering is suspended. The higher the
value, the lower is the effect due to noise and
mechanical vibration and hence having a more
accurate display.
0
(
NONE
)
1
(
2D
)
2
(
5D
)
√
3
(
10D
)
4
(
20D
)
5
(
50D
)
6
(
100D
)
F1.11
(
Digital filter
cutout threshold
)
7
(
200D
)
Digital Filter Cutout Threshold. Specifies the filter
threshold, in display divisions. When a specified
number of consecutive scale readings (the value set
for F1.10) fall outside of this threshold, digital
filtering is suspended. Digital filtering continues to
function when F1.11 is set to 0 (NONE).
0
(
Yes
)
√
F1.12
(
Power Up Zero
)
1
(
No
)
Power Up Zero. Specifies whether to zero the
scale when switching on the scale. When selected 0
(Yes), indicator zeros the scale after finishing self
checking.