3
(
FS
)
scale is overloaded.
F1.6
(
Sample
Rate
)
0
(
7.5Hz
)
Sample rate.
Selects the
measurement rate in samples
per second of the analogue-
to-digital converter. Lower
sample rate values provide
greater signal noise immunity.
1
(
15Hz
)
√
2
(
30Hz
)
3
(
60Hz
)
4
(
120Hz
)
5
(
240Hz
)
6
(
480Hz
)
7
(
960Hz
)
F1.7
(
Digital
Filter
)
0
Digital Filter
. Selects the
digital filtering rate
.
The
higher the value, the lower is
the effects of motion impact
the indicator. This results in
having a more accurate
display. However, it slows
down the settling rate of the
indicator.
Power Up Auto Zero
.
Specifies to automatically
zero the scale when switching
on the scale. When selected 0
(Yes), indicator zeros the scale
after finishing self checking.
Power Up Manual Zero
.
Specifies to restore manually
zero the scale when switching
it on. When 0 (Yes) is
selected
,
press zero button in
the indicator to zero the scale
and saves current zero for
next powering up.
1
√
2
3
4
5
6
7
F1.8
(
Power Up
Auto Zero
for
805TS-B-
17,
805BS-B-17 &
805BS-TN-B
)
0
(
Yes
)
√
1
(
No
)
F1.9
(
Power Up
Manual Zero
for
805TS-B-
17,
805BS-B-17 &
805BS-TN-B
)
0 (Yes)
√
1 (No)
Note: Parameters that are in check are the default parameter values.