9 Putting HTK 1200N out of Operation
B40IB006EN-D
39
9 Putting HTK 1200N out of Operation
When you stop using HTK 1200N for a short period
without removing it from the diffractometer, proceed
as follows:
1. Cool down the sample holder to room tempera-
ture.
2. Turn off the heater with the HEATER button.
3. Remove the sample and put the sample holder
back into the sample chamber.
4. Turn off the cooling water.
5. Turn off CCU 1000 completely.
In order to remove HTK 1200N from the diffractom-
eter continue as follows:
1. Completely turn off CCU 1000 (mains switch).
2. Disconnect all electrical cables from the sample
chamber.
3. Disconnect the couplings of the cooling water
hoses.
4. Dismount the sample chamber from the goniom-
eter.
5. Store the sample chamber in a safe place, pref-
erably on a stage mount offered by Anton Paar
GmbH.
Содержание HTK 1200N
Страница 2: ......
Страница 6: ...Contents 6 B40IB006EN D ...
Страница 56: ...Appendix G Declaration of Conformity 56 B40IB006EN D Appendix G Declaration of Conformity ...