Appendix C: Thermal Expansion of the Sample Holder
B40IB006EN-D
51
Appendix C: Thermal Expansion of the Sample
Holder
Due to the thermal expansion of the used materials
the height of the sample holder expands with in-
creasing temperature which makes regular realign-
ments during high- temperature XRD investigations
necessary.
The tables and diagrams below show measure-
ments of vertical expansion of the rigid sample hold-
er in vacuum and air. The indicated values are
comparable for the rigid sample holder and for the
sample spinner and are supposed to give a rough
idea about the magnitude of vertical expansion.
Vertical sample holder expansion
in air
:
Vertical sample holder expansion
in vacuum
:
Temperature [C°]
Vertical expansion [mm]
25
0,00
100
0,02
200
0,06
300
0,09
400
0,13
500
0,18
600
0,22
700
0,27
800
0,32
900
0,36
1000
0,41
1100
0,45
1200
0,49
Temperature [C°]
Vertical expansion [mm]
25
0,00
100
0,02
200
0,06
300
0,11
400
0,15
500
0,20
600
0,25
700
0,30
800
0,35
900
0,40
1000
0,45
1100
0,50
1200
0,54