6 Operating CCU 1000
30
B40IB006EN-D
6.4 Turning the Heater On and Off
Pressing the heater button turns on the heating
power supply to the heater inside the sample cham-
ber. The red and the green HEATER LED lights up
and the sample holder is heated to the temperature
set-point SP with the default heating rate. Changing
of the heating rate is described in the previous chap-
ter.
If the heater button is pushed again, heating is
stopped. Temperature set-point SP and ramp rate
SPR are reset to the default values. The green
HEATER LED is switched off.
TIP: After pressing the heater button it takes a short
time until a reaction of the CCU 1000 is visible.
6.5 Manual Control of CCU 1000
Manual operation of HTK 1200N is done with the
keys on the front panel of CCU 1000. Proceed as
follows to heat or cool the sample to the desired
temperature:
1. Make sure that the heater is switched on.
2. Set the desired heating/cooling rate on the
parameter page of CCU 1000. (See chapter
6.3.1.)
3. Push the scroll button one time (SP LED green)
and set the desired target temperature by using
the UP/DOWN buttons.
4. Wait until the displayed sample holder tempera-
ture has reached the target value.
TIP: If you have chosen a large heating/cooling
rate, wait 1 minute before you start the X-ray scan to
make sure the sample surface has reached a con-
stant temperature.
6.6 Remote Control of CCU 1000
Usually CCU 1000 is integrated in the software that
controls the X-ray diffractometer. For more informa-
tion about controlling HTK 1200N, read the instruc-
tion manual of the diffractometer control software.
Alternatively it is also possible to remote control the
instrument with the delivered Nambicon software
(Non-AMBIent-CONtrol) by Anton Paar GmbH. For
further information see the corresponding manual of
the Nambicon software.
The NAMBICON software is included in the delivery
of the instrument. Please follow the instructions in
the manual of the software for corresponding instal-
lation and use.
WARNING
Hot surface.
There can be dangerously hot surfaces inside
the sample chamber even if the heater has been
switched off. Always check on the display of
CCU 1000 that the sample holder temperature is
below 50 °C before you open the sample cham-
ber and touch any parts inside.