4 Installation
B40IB006EN-D
21
4.6 Installing the Cooling Water Circuit
The water flows through the HTK 1200N housing
and the lid.
The HTK 1200N is supplied with mounted cooling
water hoses with quick couplings. Water inlet (1)
and oulet (2) are located on the HTK 1200N lid.
Make sure that the water inlet is connected to nozzle
1 and the water outlet to nozzle 2!
Take care that the hoses connected to the nozzles
are always protected against unintentional discon-
nection. Appropriate nuts (3) are supplied with the
instrument.
The Water Flow Controller (part N° 164877) con-
tains a pressure regulator and a flow control unit.
The required flow rate is approx. 0.7 l/min at a pres-
sure of 2 bar. The flow controller supplied by Anton
Paar GmbH is set to this value. Refer to
for detailed information on the requirements for
cooling water quality.
Connect the cable of the flow control unit to the WA-
TER FLOW CONTROL connector on the rear panel
of CCU 1000 (Fig. 10 (5)).
4.7 Installing the Gas Supply
Depending on the specific measuring problem, the
HTK 1200N can be operated under different gas at-
mospheres, preferably under inert gas.
The gas supply tubing is connected to the Swagelok
fitting in the lid, see
The gas exhaust tubing can be connected to the
adapter. The type of gas connector depends on the
adapter type.
TIP: Refer to Appendix F for further information.
4.8 Installing the Vacuum Equipment
The HTK 1200N can be operated under a vacuum
of approx. 10
-4
mbar.
TIP: Anton Paar GmbH offers a suitable vacuum
equipment for the HTK 1200N. All required compo-
nents for the installation are included in the delivery.
Please refer to Appendix E.
Vacuum can be produced either through the adapt-
er axis in the base plate or through the vacuum
flange in the lid (see below).
NOTICE
Take care that all hoses are secured on the noz-
zles against unintentional disconnection. Appro-
priate nuts are supplied with the instrument.
NOTICE
Risk of damage
Exceeding the maximum permissible pressure of
4 bar may result in bursting of the hoses.
1
2
NOTICE
Risk of damage
-
In order to avoid damage, do not operate
the HTK 1200N before the cooling water
circuit is started.
-
Check the proper functioning of the flow
controller regularly!
NOTICE
Risk of Damage
Check the chemical resistance of the materials
used in the HTK 1200N BEFORE you start an
experiment with a gas other than specified in
WARNING
Risk of injury
Do NOT use poisonous, explosive or aggressive
gases.
Make sure that all fittings are tight and mounted
properly, and an appropriate exhaust hose is
connected to the gas outlet.