42 User manual OR 100 /110, ed. 5
C H A P T E R 5
Trouble shooting guide
All ALEXYS system components are off
Possible cause
Remedy
No power
Plug in power cord
Power switch off
Turn power switch on
Faulty fuses
Replace fuses
Divergent mains voltage
Check line voltage, see page 13
Degasser LEDs are off
Possible cause
Remedy
No power
Plug in power cord
Power switch off
Turn power switch on
Faulty OR 100 mains fuses
Replace fuses
Divergent mains voltage
Check line voltage, see page 13
Degasser yellow status LED is on or flashing
Possible cause
Remedy
Yellow LED is on
Typical normal operation (see
‘Powering up the degasser’ page 36)
Yellow LED is flashing
System leak or fault. Contact your lo-
cal supplier for service
Air bubbles in FEP/PTFE tubing
Possible cause
Remedy
Loose fittings
Check and if needed tighten the input and
output 1/8” fittings on the solvent ports
Mixing of solvents requires
more capacity of degasser
Connect both channels to double capacity
Flow rate too high
Connect both channels to double capacity
Degasser not functioning
Switch on degasser
Damaged PTFE tubing
Replace FEP/PTFE tubing