CHAPTER 2 Installation guide
21
Fig. 8. ALEXYS system with two pumps and AS 100 (top) or AS 110
(bottom) autosampler.
5. In case the OR 110 was purchased as a part of an ALEXYS system
please refer to the LC-connection kit install guide supplied with the
ALEXYS system for further instructions. If the OR 110 was purchased
stand-alone (so without ALEXYS sytem) use orange-striped PEEK
tubing (1/16” OD, 500 um ID) to make the LC connections on the high
pressure side (so from pump outlet
→
pulse damper and pulse damper
→
injector)
. Use FEP tubing (1/8” OD, 1/16” ID) to make the connec-
tion on the low pressure side (so from MP bottle
→
degasser and de-
gasser
→
pump inlet). Only use the supplied nut/ferrule set (part
180.1550/180.1552) provided in the ship kit to connect FEP tubing to
degasser. For the connections on the high pressure side standard 10-
32 PEEK Fingertights can be used.