Specifications (analog output) HR-13*1
Output rate (to be switched by
indication resolution)
1 mV/°C
(1°C resolution)
10 mV/°C
(0.1°C resolution)
Output range
Full range @1°C
resolution
measurements
Full range @0.1°C
resolution
measurements
Analog output conversion accuracy
(under the
25°C±10°C environment)
Add equivalent to
±1°C (1 mV) to
measurement
accuracy
Add equivalent to
±0.1°C (1 mV) to
measurement accuracy
Temperature coefficient
(When exceeding 25°C±10°C,
exceeded temperature Δ
t
°C is
multiplied by the coefficient and then
added to the analog output
conversion accuracy)
± (0.1 mV/°C ×Δ
t
°C)
Ex. @50°C or 0°C
environment: ± 1.5
mV (equivalent to
2°C)
± (0.1 mV/°C × Δ
t
°C)
Ex. @50°C or 0°C
environment: ± 1.5 mV
(equivalent to 0.2°C)
When the probe disconnected
About -2.3 V
Out of measurement range (+)
About 5.1 V
Out of measurement range (-)
About -2.2 V
Insulation
Insulation resistance
100 MΩ/500 VDC
(Between probe input and analog output)
Rated working voltage
300Vp-p
(Between probe input and analog output)
Cable
ANGHA-1.5
Warning
To prevent electric shock, if the probe is still in contact with the voltage application
part even when the power is off, disconnect the probe before setting the analog
output cable and equipment.
Do not touch the metal part or terminal part of the probe cable or analog output
cable during measurement.
* When measuring under the influence of high voltage and high frequency, there is a
risk of electric shock and measurement failure, so please contact us.
* The probe input and analog output are isolated, but they are connected by a capacitor
to reduce output noise.
(Leakage current 0.5mA or less in AC500V withstand voltage test)