15
6. Analog Output
[Applicable Model: HR-13*1]
Analog output has two output rates. Press the RESO key to switch the output
rates. When the analog output segment is illuminating/flashing, the output is
always active.
Output rate
Output range
1 mV/°C
All measurement range with a
1
°C
resolution
10 mV/°C
Measurement range of -100 to
500
°C
with a 0.1
°C
resolution
Note: No automatic output rate switching with the temperature range
In overrange or when the sensor element is broken (burnout), the analog output
is fixed as follows:
Higher than the upper limit: About 5,100 mV
Lower than the lower limit: About
−
2,200 mV
Burnout: About
−
2,300 mV
Signals and pin assignments
Pin #
Signal
5
+
6
-
(Dedicated cable: ANGHA-1.5)
How to connect the recorder
(1)
Turn the power of the recorder and the instrument OFF.
(2)
Connect the red terminal of the analog output cable to the [+] terminal of
the recorder and the black terminal to the [-] terminal of the recorder.
(3)
Align the input range of the recorder with the measurement range of the
instrument.
(4)
Turn the power of the recorder and the instrument ON.
(5)
Make sure that the Automatic Power OFF function is released. If not,
press the AUTO OFF key once to release it.
Illuminating
:
10 mV/
C
Flashing
:
1 mV/
C