A-30
690XXB/691XXB MM
3-11 Power Level Accuracy And Flatness Tests (Continued)
(Model 69037B/69137B w/Option 15A High Power & w/Option 2A Step Attenuator)
Power Level Accuracy Test Procedure
Set F1 to 5.0 GHz
Set L1 to:
Measured Power *
+15 dBm
+14 dBm
+13 dBm
+12 dBm
+11 dBm
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is
±
1.0 dB.
Power Level Flatness Test Procedure (Manual Sweep)
Set L1 to:
Max Power
Min Power
Variation **
+15 dBm
_________dBm
__________dBm
_________dB
** Maximum variation is 1.6 dB.
Power Level Flatness Test Procedure (Analog Sweep) (Model 69137B only)
Set L1 to:
Max Power
Min Power
Variation ***
+15 dBm
_________dBm
__________dBm
_________dB
*** Maximum variation is 6.0 dB (typical, not a specification).