Alerts, Warnings and Alarms
Alert: Low BG
F
Cause
BG entry below 70 mg/dL.
Effect
Requires user confirmation to continue.
Message
Displayed until confirmed.
Action
Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: High BG
F
Cause
BG entry above 250 mg/dL.
Effect
Requires user confirmation to continue.
Message
Displayed until confirmed.
Action
Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: Clearing Basal Segments
Cause
Clear command selected from Basal Options screen.
Effect
Requires user confirmation to continue.
Message
Displayed until confirmed or until pump goes to sleep.
Action
Select Clear Program or Basal Options.
Beeps/Vib
User selected, one time.
Alert: Suspend (Temp Basal/Combo Bolus Canceled)
Cause Pump
suspended.
Effect Any
active
Temp
Basal/Combo Bolus canceled.
Message
Displayed once for 3 seconds.
Action None
required.
Beeps/Vib User selected, one time. No progression.
Содержание IR 1250
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 20: ...Page 20...
Страница 44: ...Page 44...
Страница 54: ...Page 54...
Страница 62: ...Page 62...
Страница 78: ...Page 78...
Страница 80: ...Page 80...
Страница 92: ...Page 92...
Страница 94: ...Page 94...
Страница 104: ...Page 104...
Страница 113: ...Page 113...