Page 38
7
Note: You do not have to Suspend the pump to edit an active program. When you select Edit
from the Basal Menu, your IR 1250 automatically suspends delivery. When you exit the Edit
mode, the active program delivery automatically resumes.
4. The
Basal Options
screen is displayed. Edit is highlighted. Press the OK button. From the
Edit Basal
screen, you can edit the basal segments of the selected program.
5. Scroll to the desired U/hr field. Press OK to activate Edit mode (indicated by flashing
cursor).
6. Use Up/Down arrows to set desired basal rate. Press OK to confirm and exit Edit mode for
this field.
7. Scroll down to select the next Start time field. Press the OK button to activate Edit mode
(indicated by flashing cursor).
Note: The next available empty basal segment will appear automatically as you program the
previous segment. If the next empty basal segment does not appear, you have programmed all
12 possible segments.
8. Change next Start time field as desired, press the OK button to exit Edit mode. Segments can
start on the hour or half hour.
Note: The 24 hour Total changes automatically as you change U/hr settings.
9. Continue until basal segments have been set as recommended by your healthcare team.
Содержание IR 1250
Страница 1: ...Page 1...
Страница 6: ...Page 6...
Страница 11: ...Page 11 Radiology Quick Reference Chart 1...
Страница 16: ...Page 16...
Страница 18: ...Page 18...
Страница 20: ...Page 20...
Страница 44: ...Page 44...
Страница 54: ...Page 54...
Страница 62: ...Page 62...
Страница 78: ...Page 78...
Страница 80: ...Page 80...
Страница 92: ...Page 92...
Страница 94: ...Page 94...
Страница 104: ...Page 104...
Страница 113: ...Page 113...