36.
Using the Emergency Stop Button
The emergency stop button is located at the foot end of the stretcher beneath the user
interface, see (A) in Fig. 17. The emergency stop button is used to deactivate all of the
powered functions in the unlikely event of a powered function auto-actuating.
The emergency stop button is activated by depressing the button (B). When activated,
the button remains depressed, and all of the LED’s on both user interfaces will flash.
None of the powered functions will now operate.
To release the emergency stop button, two actions are required. Firstly, the button is
rotated clockwise to release it (C), and secondly, the ON/OFF button on the user
interface must be pressed.
37.
X-ray Cassette Detection Sensor
See Section 17. ‘Additional Features of Emergency Stretcher’.
Powered Emergency stretchers are fitted with a sensor to detect if there is an x-ray
cassette, or digital plate, positioned across the backrest hinge. If an x-ray cassette or
digital plate is detected, the sensor will disable the powered backrest from raising or
lowering, the ‘READ INSTRUCTIONS’ indicator will flash amber and a beep will sound.
CAUTION: Do not raise the backrest with the x-ray tray, x-ray cassette or digital
plate across the backrest hinge as damage may occur.
38.
Battery Charge Level Indication, Battery Charging & Maintenance
See Section 24. ‘Battery Charge Level Indication, Battery Charging & Maintenance’.
Содержание QA3 21110
Страница 1: ...Document No 992012 Issue 19 24 10 22...
Страница 2: ...Document No 992012 Issue 19 24 10 22...
Страница 8: ...Product Specifications Document No 992012 Issue 19 24 10 22 Page 4 2 Product Specifications Fig 1...
Страница 36: ...Product Accessories Document No 992012 Issue 19 24 10 22 Page 32...
Страница 40: ...Document No 992012 Issue 19 24 10 22 Page 36...