a.  After  applying  the  cuff  and  your  body  is  in  a  comfortable  position,  press  the 

“START” button. A beep is heard and all display characters are shown for self-test. 

See picture 7.

 

Please contact the service center if segment is missing. 

b.  Then  the  current  memory  bank  (U1  or  U2)  is  blinking. 

See  picture  7-1.

  Press 

“MEM”  button  to  change  over  to  other  bank.

  See  picture  7-2.

  Confirm  your 

selection  by  pressing  “START”  button.  The  current  bank  can  also  be  confirmed 

automatically after 5 seconds with no operation. 

         
 
 
 
 
 
         

Picture 7                              Picture 7-1                        Picture 7-2 

c.  After  selecting  the  memory  bank,  the  monitor  starts  to  seek  zero  pressure.

  See 

picture 7-3.

 

d. 

Then  the  cuff  will  be  slowly  inflated.  The  blood  pressure  and  pulse  will  be 

measured  during  inflation.  Inflation  will  stop  as  soon  as  the  blood  pressure  and 

pulse  rate  have  been  calculated  and  displayed  on  the  LCD.  The  irregular 

heartbeat symbol (if any) and blood pressure classification indicator will blink on 

the LCD,

 

See picture 7-4

. The result will automatically be stored in the Memory 

bank of the monitor

 

 

 
 
 
 

Picture 7-3                        Picture 7-4

 

e.  After  measurement,  the  monitor  will  turn  off  automatically  after  1  minute  of  no 

operation. Alternatively, you can press the “START” button to turn off the monitor 

manually. 

f. 

During measurement, you can press the “START” button to turn off the monitor 

manually. 

Note: Please 

consult a health care professional for interpretation of pressure measurements.

 

 

 

8.  DISPLAYING STORED RESULTS 

 

a. 

In StandBy 

Mode, press “MEM” button, the monitor will blink sign of current group. 

See picture 8.

 

 

 
 
 

 

Содержание KD-723

Страница 1: ...OCEDURES 5 1 DOWNLOAD THE FREE APP 5 2 BATTERY LOADING 5 3 CLOCK AND DATE ADJUSTMENT 5 4 CONNECTING THE CUFF TO THE MONITOR 6 5 APPLYING THE CUFF 7 6 BODY POSTURE DURING MEASUREMENT 7 7 TAKING YOUR BL...

Страница 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Страница 3: ...ored in the memory with date and time stamp The voice function will ease the operation The Electronic Sphygmomanometers corresponds to the below standards IEC 60601 1 2005 corr 1 2006 corr 2 2007 EN 6...

Страница 4: ...0 4 122 15 Environmental humidity for storage and transport 85 RH 16 Environmental pressure 80kPa 105kPa 17 Battery life Approx 170 times 18 Wireless Connection Smart Bluetooth Frequency Band 2 400 2...

Страница 5: ...formation regarding potential electromagnetic or other interference between the blood pressure monitor and other devices together with advice regarding avoidance of such interference please see part E...

Страница 6: ...f according to local regulations at the end of their usage 3 CLOCK AND DATE ADJUSTMENT a At first the Blood Pressure Monitor is totally off once you insert the battery the Blood Pressure Monitor will...

Страница 7: ...or by pressing START STOP button when the minute is blinking then the time and date is confirmed Note 3 1 The clock format could be set by user 4 CONNECTING THE CUFF TO THE MONITOR The cuff is attache...

Страница 8: ...o not move your arm body or the monitor during measurement Stay still calm for 5 minutes before blood pressure measurement Please keep the cuff clean Clean the cuff by wet soft cloth and mild detergen...

Страница 9: ...ly inflated The blood pressure and pulse will be measured during inflation Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on the LCD The irregular...

Страница 10: ...Press MEM button again to review the next result See picture 8 6 In this way repeatedly pressing the MEM button displays the respective results measured previously Picture 8 5 Picture 8 6 e When disp...

Страница 11: ...e different levels of blood pressure 12 TECHNICAL ALARM DESCRIPTION The monitor will show HI or Lo as technical alarm on LCD with no delay if the determined blood pressure systolic or diastolic is out...

Страница 12: ...to detect diastolic pressure LCD shows Er 3 Pneumatic system blocked or cuff is too tight during inflation Apply the cuff correctly and try again LCD shows Er 4 Pneumatic system leakage or cuff is to...

Страница 13: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Страница 14: ...ANT INFORMATION REQUIRED BY THE R TTE This product is approved in accordance to R TTE directive transmitter interference This product complies with Industry Canada IC RSS 247 The Bluetooth word mark a...

Страница 15: ...e KD 723 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 723 should assure that it is used in such an environment IMMUNITY test IEC 60601test...

Страница 16: ...NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from struct...

Страница 17: ...aximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz...

Страница 18: ...rgy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particu...

Отзывы: